Wall shear rate (WSR) is an important stimulus for the brachial artery flow-mediated dilation (FMD) response. However, WSR estimation near the arterial wall by conventional Doppler is inherently difficult. To overcome this limitation, we utilized multigate Doppler to accurately determine the WSR stimulus near the vessel wall simultaneously with the FMD response using an integrated FMD system [Ultrasound Advanced Open Platform (ULA-OP)]. Using the system, we aimed to perform a detailed analysis of WSR-FMD response and establish novel WSR parameters in a healthy young population. Data from 33 young healthy individuals (27.5 ± 4.9 yr, 19 females) were analyzed. FMD was assessed with reactive hyperemia using ULAOP. All acquired raw data were postprocessed using customdesigned software to obtain WSR and diameter parameters. The acquired velocity data revealed that nonparabolic flow profiles within the cardiac cycle and under different flow states, with heterogeneity between participants. We also identified seven WSR magnitude and four WSR time-course parameters. Among them, WSR area under the curve until its return to baseline was the strongest predictor of the absolute (R2 = 0.25) and percent (R2 = 0.31) diameter changes in response to reactive hyperemia. For the first time, we identified monoand biphasic WSR stimulus patterns within our cohort that produced different magnitudes of FMD response [absolute diameter change: 0.24 ± 0.10 mm (monophasic) vs. 0.17 ± 0.09 mm (biphasic), P < 0.05]. We concluded that accurate and detailed measurement of the WSR stimulus is important to comprehensively understand the FMD response and that this advance in current FMD technology could be important to better understand vascular physiology and pathology. Copyright © 2018 American Physiological Society. All rights reserved.

Brachial artery vasodilatory response and wall shear rate determined by multi-gate Doppler in a healthy young cohort

Morizzo C
Investigation
;
Palombo C.
Ultimo
Conceptualization
2018-01-01

Abstract

Wall shear rate (WSR) is an important stimulus for the brachial artery flow-mediated dilation (FMD) response. However, WSR estimation near the arterial wall by conventional Doppler is inherently difficult. To overcome this limitation, we utilized multigate Doppler to accurately determine the WSR stimulus near the vessel wall simultaneously with the FMD response using an integrated FMD system [Ultrasound Advanced Open Platform (ULA-OP)]. Using the system, we aimed to perform a detailed analysis of WSR-FMD response and establish novel WSR parameters in a healthy young population. Data from 33 young healthy individuals (27.5 ± 4.9 yr, 19 females) were analyzed. FMD was assessed with reactive hyperemia using ULAOP. All acquired raw data were postprocessed using customdesigned software to obtain WSR and diameter parameters. The acquired velocity data revealed that nonparabolic flow profiles within the cardiac cycle and under different flow states, with heterogeneity between participants. We also identified seven WSR magnitude and four WSR time-course parameters. Among them, WSR area under the curve until its return to baseline was the strongest predictor of the absolute (R2 = 0.25) and percent (R2 = 0.31) diameter changes in response to reactive hyperemia. For the first time, we identified monoand biphasic WSR stimulus patterns within our cohort that produced different magnitudes of FMD response [absolute diameter change: 0.24 ± 0.10 mm (monophasic) vs. 0.17 ± 0.09 mm (biphasic), P < 0.05]. We concluded that accurate and detailed measurement of the WSR stimulus is important to comprehensively understand the FMD response and that this advance in current FMD technology could be important to better understand vascular physiology and pathology. Copyright © 2018 American Physiological Society. All rights reserved.
2018
Aizawa, K; Sbragi, S; Ramalli, A; Tortoli, P; Casanova, F; Morizzo, C; Thorn, C; Shore, A; Gates, P; Palombo, C.
File in questo prodotto:
File Dimensione Formato  
ULA-OP FMD healthy young paper 2nd revised main text for JAP ver 3.0.docx

accesso aperto

Descrizione: manoscritto accettato
Tipologia: Documento in Pre-print
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 95.21 kB
Formato Microsoft Word XML
95.21 kB Microsoft Word XML Visualizza/Apri
japplphysiol.2018.pdf

solo utenti autorizzati

Descrizione: Lavoro in extenso
Tipologia: Versione finale editoriale
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 2.82 MB
Formato Adobe PDF
2.82 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/872560
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
social impact