Background: The escalating prevalence of healthcare-associated infections (HAIs) driven by multidrug-resistant (MDR) pathogens presents a critical challenge in contemporary veterinary medicine. The reprocessing of surgical instruments and complex medical devices requires rigorous, standardized protocols to prevent cross-contamination and subsequent nosocomial outbreaks. This study evaluates the antimicrobial efficacy of a novel sterilization system combining gaseous Ozone (O3) and Ultraviolet-C (UV-C) radiation against MDR Klebsiella pneumoniae and S. pseudintermedius. Methods: Two distinct equipment sets were artificially contaminated with an MDR K. pneumoniae suspension (Set 1; 1.5 x 108 CFU/mL) and MDR S. pseudintermedius (Set 2; 1.5 x 108 CFU/mL), and allowed to form dry surface biofilms. Set 1 comprised 10 stainless-steel surgical instruments with complex mechanical joints (box locks and ratchets). Set 2 comprised 10 heterogeneous devices characterized by complex geometries, internal lumens, and porous materials (laryngoscopes, clipper blades, endotracheal tubes, high-flow nasal cannulas, and non-invasive blood pressure cuffs). Three pre-treatment protocols were evaluated before Ozone/UV-C exposure: Protocol A (direct packaging, simulating non-compliant omission of pre-treatment), Protocol B (standard decontamination and mechanical cleaning), and Protocol C (complete protocol including mechanical cleaning and high-level chemical disinfection). Efficacy was evaluated via qualitative swabbing for nominal data, and quantitative colony-forming unit (CFU) counting for continuous data. Results: Under the non-compliant Protocol A, sterilization failed in 40% of the standard surgical instruments due to bacterial survival within mechanical joints. A 100% failure rate was recorded for structurally complex devices, which exhibited significant bacterial proliferation across all porous and luminal geometries (median >100 CFU). Conversely, Protocols B and C achieved complete sterilization and total bacterial eradication (0 CFU). The reduction in bacterial load between Protocol A and Protocols B/C was statistically significant across both datasets (Friedman test, p < 0.001). No statistically significant difference in antimicrobial efficacy was observed between Protocol B and Protocol C. Conclusions: The synergistic application of Ozone and UV-C technology demonstrates exceptionally high bactericidal efficacy against MDR K. pneumoniae and S. pseudintermedius. However, organic bioburden acts as a critical physical shield against UV-C radiation and Ozone diffusion. Strict adherence to mechanical cleaning and standard decontamination is mandatory to ensure the device’s sterilisation efficacy, particularly for instruments with complex joints, internal lumens, and porous matrices.
Efficacy of an Ozone and UV-C Sterilization System on Veterinary Surgical Instruments and Complex Medical Devices Contaminated with MDR Klebsiella pneumoniae and Staphylococcus pseudintermedius
Matilde Vernaccini
Primo
;Ilaria Lippi;Angela BrigantiUltimo
2026-01-01
Abstract
Background: The escalating prevalence of healthcare-associated infections (HAIs) driven by multidrug-resistant (MDR) pathogens presents a critical challenge in contemporary veterinary medicine. The reprocessing of surgical instruments and complex medical devices requires rigorous, standardized protocols to prevent cross-contamination and subsequent nosocomial outbreaks. This study evaluates the antimicrobial efficacy of a novel sterilization system combining gaseous Ozone (O3) and Ultraviolet-C (UV-C) radiation against MDR Klebsiella pneumoniae and S. pseudintermedius. Methods: Two distinct equipment sets were artificially contaminated with an MDR K. pneumoniae suspension (Set 1; 1.5 x 108 CFU/mL) and MDR S. pseudintermedius (Set 2; 1.5 x 108 CFU/mL), and allowed to form dry surface biofilms. Set 1 comprised 10 stainless-steel surgical instruments with complex mechanical joints (box locks and ratchets). Set 2 comprised 10 heterogeneous devices characterized by complex geometries, internal lumens, and porous materials (laryngoscopes, clipper blades, endotracheal tubes, high-flow nasal cannulas, and non-invasive blood pressure cuffs). Three pre-treatment protocols were evaluated before Ozone/UV-C exposure: Protocol A (direct packaging, simulating non-compliant omission of pre-treatment), Protocol B (standard decontamination and mechanical cleaning), and Protocol C (complete protocol including mechanical cleaning and high-level chemical disinfection). Efficacy was evaluated via qualitative swabbing for nominal data, and quantitative colony-forming unit (CFU) counting for continuous data. Results: Under the non-compliant Protocol A, sterilization failed in 40% of the standard surgical instruments due to bacterial survival within mechanical joints. A 100% failure rate was recorded for structurally complex devices, which exhibited significant bacterial proliferation across all porous and luminal geometries (median >100 CFU). Conversely, Protocols B and C achieved complete sterilization and total bacterial eradication (0 CFU). The reduction in bacterial load between Protocol A and Protocols B/C was statistically significant across both datasets (Friedman test, p < 0.001). No statistically significant difference in antimicrobial efficacy was observed between Protocol B and Protocol C. Conclusions: The synergistic application of Ozone and UV-C technology demonstrates exceptionally high bactericidal efficacy against MDR K. pneumoniae and S. pseudintermedius. However, organic bioburden acts as a critical physical shield against UV-C radiation and Ozone diffusion. Strict adherence to mechanical cleaning and standard decontamination is mandatory to ensure the device’s sterilisation efficacy, particularly for instruments with complex joints, internal lumens, and porous matrices.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.