This article describes the fabrication and characterization of a pH sensor for monitoring the wound status. The pH sensitive layer consists of a graphene oxide (GO) layer obtained by drop-casting 5 μl of GO dispersion onto the working electrode of a screen-printed substrate. Sensitivity was 31.8 mV/pH with an accuracy of 0.3 unit of pH. Open-circuit potentiometry was carried out to measure pH in an exudate sample. The GO pH sensor proved to be reliable as the comparison with results obtained from a standard glass electrode pH-meter showed negligible differences (< 0.09 pH units in the worst case) for measurements performed over a period of 4 days.
A graphene oxide pH sensor for wound monitoring
MELAI, BERNARDO;SALVO, PIETRO;CALISI, NICOLA;Bonini, A.;PAOLETTI, CLARA;LOMONACO, TOMMASO;MOLLICA, VINCENZO;FUOCO, ROGER;DI FRANCESCO, FABIO
2016-01-01
Abstract
This article describes the fabrication and characterization of a pH sensor for monitoring the wound status. The pH sensitive layer consists of a graphene oxide (GO) layer obtained by drop-casting 5 μl of GO dispersion onto the working electrode of a screen-printed substrate. Sensitivity was 31.8 mV/pH with an accuracy of 0.3 unit of pH. Open-circuit potentiometry was carried out to measure pH in an exudate sample. The GO pH sensor proved to be reliable as the comparison with results obtained from a standard glass electrode pH-meter showed negligible differences (< 0.09 pH units in the worst case) for measurements performed over a period of 4 days.File | Dimensione | Formato | |
---|---|---|---|
A graphene oxide pH sensor for wound monitoring.pdf
solo utenti autorizzati
Descrizione: Proceeding Conferenza
Tipologia:
Versione finale editoriale
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
698.54 kB
Formato
Adobe PDF
|
698.54 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.