Transapical off-pump mitral valve repair with neochord implantation is an established technique for minimally-invasive intervention on mitral valve prolapse/flail. The procedure involves the positioning of artificial chords, whose length/tension is adjusted intraoperatively, adopting different methods based on the experience of the surgeon. This unsystematic approach occasionally leads to complications such as leaflet rupture and excessive/insufficient load on the neochords. In this study, finite element models of a generalized prolapsing mitral valve are used to verify the effect of two alternative tensioning approaches (AT – All together and 1-by-1 – one by one sequences) on the coaptation area and valve biomechanics, comparing results with a corresponding healthy configuration. The total force of about 1 N is exerted by the chords in both strategies, but the maximum stress and coaptation area are closer to those of the healthy configuration in the 1-by-1 sequence. However, the analysis also provides an explanation for the chords unloading in the 1-by-1 strategy observed in the clinical practice, and suggests an optimum tensioning methodology for NeoChord procedures. The study also reveals the potential power of the implemented numerical approach to serve as a tool for procedural planning, supporting the identification of the most suitable ventricular access site and the most effective stitching points for the artificial chords.

The neochord mitral valve repair procedure: Numerical simulation of different neochords tensioning protocols

Colli A.
Conceptualization
;
Besola L.;
2019-01-01

Abstract

Transapical off-pump mitral valve repair with neochord implantation is an established technique for minimally-invasive intervention on mitral valve prolapse/flail. The procedure involves the positioning of artificial chords, whose length/tension is adjusted intraoperatively, adopting different methods based on the experience of the surgeon. This unsystematic approach occasionally leads to complications such as leaflet rupture and excessive/insufficient load on the neochords. In this study, finite element models of a generalized prolapsing mitral valve are used to verify the effect of two alternative tensioning approaches (AT – All together and 1-by-1 – one by one sequences) on the coaptation area and valve biomechanics, comparing results with a corresponding healthy configuration. The total force of about 1 N is exerted by the chords in both strategies, but the maximum stress and coaptation area are closer to those of the healthy configuration in the 1-by-1 sequence. However, the analysis also provides an explanation for the chords unloading in the 1-by-1 strategy observed in the clinical practice, and suggests an optimum tensioning methodology for NeoChord procedures. The study also reveals the potential power of the implemented numerical approach to serve as a tool for procedural planning, supporting the identification of the most suitable ventricular access site and the most effective stitching points for the artificial chords.
2019
Di Micco, L.; Peruzzo, P.; Colli, A.; Burriesci, G.; Boso, D.; Besola, L.; Gerosa, G.; Susin, F. M.
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S1350453319301900-main.pdf

non disponibili

Tipologia: Versione finale editoriale
Licenza: NON PUBBLICO - accesso privato/ristretto
Dimensione 2.22 MB
Formato Adobe PDF
2.22 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Accepted version.pdf

accesso aperto

Tipologia: Documento in Post-print
Licenza: Creative commons
Dimensione 3.41 MB
Formato Adobe PDF
3.41 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/1025990
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
social impact