BackgroundThe Najuta stent-graft (Kawasumi Laboratories Inc., Tokyo, Japan) is usually easily advanced to the correct deployment position in the ascending aorta thanks to the pre-curved delivery J-sheath with all fenestrations automatically oriented towards the supra-aortic vessels. Aortic arch anatomy and delivery system stiffness could however represent limitations for proper endograft advancement, especially when the aortic arch bends sharply. The aim of this technical note is to report a series of bail-out procedures that could be useful to overcome the difficulties encountered during the Najuta stent-graft advancement up to the ascending aorta.Main bodyThe insertion, positioning and deployment of a Najuta stent-graft requires a through-and-through guidewire technique using a .035 '' 400 cm hydrophilic nitinol guidewire (Radifocus (TM) Guidewire M Non-Vascular, Terumo Corporation, Tokyo, Japan) with right brachial and both femoral accesses. When standard maneuver to put the endograft tip into the aortic arch, some bail-out procedures can be applied to obtain proper positioning. Five techniques are described into the text: positioning of a coaxial extra-stiff guidewire; positioning of a long introducer sheath down to the aortic root from the right brachial access; inflation of a balloon inside the ostia of the supra-aortic vessels; inflation of a balloon inside the aortic arch (coaxial to the device); and transapical access technique. This is a troubleshooting guide for allowing physicians to overcome various difficulties with the Najuta endograft as well as for other similar devices.Short conclusionTechnical issues in advancing the delivery system of Najuta stent-graft could occur. Therefore, the rescue procedures described in this technical note could be useful to guarantee the correct positioning and deployment of the stent-graft.

A technical review of bail-out procedures to place Najuta stent-graft into the ascending aorta

Berchiolli, Raffaella
Primo
Writing – Review & Editing
;
Troisi, Nicola
Secondo
Writing – Original Draft Preparation
;
Bertagna, Giulia;Colli, Andrea;
2023-01-01

Abstract

BackgroundThe Najuta stent-graft (Kawasumi Laboratories Inc., Tokyo, Japan) is usually easily advanced to the correct deployment position in the ascending aorta thanks to the pre-curved delivery J-sheath with all fenestrations automatically oriented towards the supra-aortic vessels. Aortic arch anatomy and delivery system stiffness could however represent limitations for proper endograft advancement, especially when the aortic arch bends sharply. The aim of this technical note is to report a series of bail-out procedures that could be useful to overcome the difficulties encountered during the Najuta stent-graft advancement up to the ascending aorta.Main bodyThe insertion, positioning and deployment of a Najuta stent-graft requires a through-and-through guidewire technique using a .035 '' 400 cm hydrophilic nitinol guidewire (Radifocus (TM) Guidewire M Non-Vascular, Terumo Corporation, Tokyo, Japan) with right brachial and both femoral accesses. When standard maneuver to put the endograft tip into the aortic arch, some bail-out procedures can be applied to obtain proper positioning. Five techniques are described into the text: positioning of a coaxial extra-stiff guidewire; positioning of a long introducer sheath down to the aortic root from the right brachial access; inflation of a balloon inside the ostia of the supra-aortic vessels; inflation of a balloon inside the aortic arch (coaxial to the device); and transapical access technique. This is a troubleshooting guide for allowing physicians to overcome various difficulties with the Najuta endograft as well as for other similar devices.Short conclusionTechnical issues in advancing the delivery system of Najuta stent-graft could occur. Therefore, the rescue procedures described in this technical note could be useful to guarantee the correct positioning and deployment of the stent-graft.
2023
Berchiolli, Raffaella; Troisi, Nicola; Bertagna, Giulia; Colli, Andrea; Besola, Laura; Silingardi, Roberto; Simonte, Gioele; Isernia, Giacomo
File in questo prodotto:
File Dimensione Formato  
Najuta_CVIR Endovascular (2).pdf

accesso aperto

Tipologia: Versione finale editoriale
Licenza: Creative commons
Dimensione 1.31 MB
Formato Adobe PDF
1.31 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/1171507
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact