D-dimer is a neoantigen formed when thrombin initiates the transformation of fibrinogen to fibrin; it is derived from plasmin digestion of cross-linked fibrin. In human medicine, the usefulness of this analyte in diagnosing disseminated intravascular coagulation (DIC) has been assessed in patients fulfilling the clini- cal and laboratory requirements for this disorder. In canine medicine, the use of D-dimer is relatively new. Detailed studies are needed to understand the relationship between D-dimer concentration in plasma and DIC status in dogs. We validated a D-dimer immunoturbidimetric assay (Tina-quant [a] D-Dimer, Boehringer Mannheim) in canine citrated plasma samples. Intra-assay and interassay variability (coefficient of variation) was 5.63% and 8.82%, respectively. The assay was linear, using 2 samples with low and high D-dimer concen- trations (r = .996 and .998). Accuracy was 102.2% and 95.7% based on a recovery study in which 2 samples were assessed. Reference values for D-dimer were established using 70 healthy dogs that were assessed clinically and evaluated on the basis of a complete laboratory workup. The reference range was set between 0.02 and 0.28 μg/mL (chi-square test for normal distribution, P > .05).
Validation of an immunoturbidimetric D-dimer assay in canine citrated plasma
LUBAS, GEORGE
2000-01-01
Abstract
D-dimer is a neoantigen formed when thrombin initiates the transformation of fibrinogen to fibrin; it is derived from plasmin digestion of cross-linked fibrin. In human medicine, the usefulness of this analyte in diagnosing disseminated intravascular coagulation (DIC) has been assessed in patients fulfilling the clini- cal and laboratory requirements for this disorder. In canine medicine, the use of D-dimer is relatively new. Detailed studies are needed to understand the relationship between D-dimer concentration in plasma and DIC status in dogs. We validated a D-dimer immunoturbidimetric assay (Tina-quant [a] D-Dimer, Boehringer Mannheim) in canine citrated plasma samples. Intra-assay and interassay variability (coefficient of variation) was 5.63% and 8.82%, respectively. The assay was linear, using 2 samples with low and high D-dimer concen- trations (r = .996 and .998). Accuracy was 102.2% and 95.7% based on a recovery study in which 2 samples were assessed. Reference values for D-dimer were established using 70 healthy dogs that were assessed clinically and evaluated on the basis of a complete laboratory workup. The reference range was set between 0.02 and 0.28 μg/mL (chi-square test for normal distribution, P > .05).I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.