A procedure is described for simultaneous determination of arsenic, selenium, tin and mercury in aqueous solution by non-dispersive atomic-fluorescence spectrometry. Radiofrequency-excited EDLs, 100% modulated in the kHz region, were used for atom excitation. Sodium tetrahydroborate was used as reductant and a hydrogen-argon miniflame as atomizer. In the optimized procedure, which uses 1 ml of sample, the limits of detection (three times the standard deviation of the blank) were 0.04, 0.08, 0.1 and 0.1 ng ml for arsenic, selenium, tin and mercury respectively. The linear dynamic range was greater than three decades for all analytes and the precision was better than 7% (typically 3%) for concentrations ≥ 1 ng ml. Results for mutual interference effects are reported. Copper, nickel, lead and cobalt interfered only with selenium (5 ng ml), when present in at least 200-fold weight ratio to it. Using 5 ml of sample improved the limits of detection for selenium and arsenic (0.01 and 0.02 ng ml respectively), but at the expense of greater interference. Recovery from spiked natural water samples was better than 95% at the ng ml level, except for selenium in sea-water, when the recovery was only 85%. Determination of the four elements, including standard-addition and background measurements, requires about 10 min.

SIMULTANEOUS DETERMINATION OF ARSENIC, SELENIUM, TIN AND MERCURY BY NON-DISPERSIVE ATOMIC FLUORESCENCE SPECTROMETRY

FUOCO, ROGER;
1985-01-01

Abstract

A procedure is described for simultaneous determination of arsenic, selenium, tin and mercury in aqueous solution by non-dispersive atomic-fluorescence spectrometry. Radiofrequency-excited EDLs, 100% modulated in the kHz region, were used for atom excitation. Sodium tetrahydroborate was used as reductant and a hydrogen-argon miniflame as atomizer. In the optimized procedure, which uses 1 ml of sample, the limits of detection (three times the standard deviation of the blank) were 0.04, 0.08, 0.1 and 0.1 ng ml for arsenic, selenium, tin and mercury respectively. The linear dynamic range was greater than three decades for all analytes and the precision was better than 7% (typically 3%) for concentrations ≥ 1 ng ml. Results for mutual interference effects are reported. Copper, nickel, lead and cobalt interfered only with selenium (5 ng ml), when present in at least 200-fold weight ratio to it. Using 5 ml of sample improved the limits of detection for selenium and arsenic (0.01 and 0.02 ng ml respectively), but at the expense of greater interference. Recovery from spiked natural water samples was better than 95% at the ng ml level, except for selenium in sea-water, when the recovery was only 85%. Determination of the four elements, including standard-addition and background measurements, requires about 10 min.
1985
Dulivo, A; Fuoco, Roger; Papoff, P.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/4143
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 23
social impact