Purpose: We report the clinical, radiological and wear analysis of 52 consecutive MoM hip resurfacings (performed on 49 younger patients) to a mean follow-up of 9.2 years. Methods: Every patient underwent X-ray and clinical evaluation (HHS). Ultrasonography of the hip was performed in all patients in order to identify possible cystic or solid mass in periprosthetic tissue. In case of mass >20 mm, further MRI was performed to better analyse the characteristics of lesion. Results: Five patients (five hips) had a revision. The overall survival rate was 90.38 %. The average HHS at follow-up examination was 95.5 points. No progressive radiolucent areas and no sclerosis or osteolysis around the implants were found. The US and RMI imaging showed a pseudotumour formation in two patients (correlated with high metal ion levels in blood and urine), both asymptomatic. Conclusion: A significant positive correlation between inclination of the acetabular component and serum metal ion levels was found (r = 0.64 and r = 0.62 for cobalt and chromium, respectively).
Metal-on-metal hip resurfacing: Correlation between clinical and radiological assessment, metalions and ultrasound findings
SCAGLIONE, MICHELANGELO;GUIDO, GIULIO
2014-01-01
Abstract
Purpose: We report the clinical, radiological and wear analysis of 52 consecutive MoM hip resurfacings (performed on 49 younger patients) to a mean follow-up of 9.2 years. Methods: Every patient underwent X-ray and clinical evaluation (HHS). Ultrasonography of the hip was performed in all patients in order to identify possible cystic or solid mass in periprosthetic tissue. In case of mass >20 mm, further MRI was performed to better analyse the characteristics of lesion. Results: Five patients (five hips) had a revision. The overall survival rate was 90.38 %. The average HHS at follow-up examination was 95.5 points. No progressive radiolucent areas and no sclerosis or osteolysis around the implants were found. The US and RMI imaging showed a pseudotumour formation in two patients (correlated with high metal ion levels in blood and urine), both asymptomatic. Conclusion: A significant positive correlation between inclination of the acetabular component and serum metal ion levels was found (r = 0.64 and r = 0.62 for cobalt and chromium, respectively).I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.