This study presents a comprehensive numerical investigation on contact mechanics of ceramic-on-ceramic hip replacements under edge-loading condition. In particular, the combined effect of implant positioning and design on the contact pressure is analysed for different load levels. The final aim is to identify design/configuration solutions which minimize the risks associated to edge-loading.

FE analyses of edge loading in ceramic-on-ceramic hip replacements with different rim design

DI PUCCIO, FRANCESCA;MATTEI, LORENZA;DI PUCCIO, FRANCESCA
2015-01-01

Abstract

This study presents a comprehensive numerical investigation on contact mechanics of ceramic-on-ceramic hip replacements under edge-loading condition. In particular, the combined effect of implant positioning and design on the contact pressure is analysed for different load levels. The final aim is to identify design/configuration solutions which minimize the risks associated to edge-loading.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.