Background No indication are available for transcatheter aortic valve implantation (TAVI) sizing in bicuspid aortic valve (BAV). Aim of the study is to develop and validate a Multi-Slice Computed Tomography (MSCT)-based algorithm for transcatheter heart valve (THV) sizing in patients with stenotic BAV under evaluation for TAVI. Methods A two steps method was applied: 1)evaluation of a cohort of 19 consecutive patients with type I BAV stenosis undergoing TAVI through pre and post-procedural MSCT, and development of an algorithm for THV sizing; 2)validation of the algorithm on a new cohort of 21 patients. Results In the first cohort, a high correlation was found between the raphe-level area measured at pre-procedural MSCT and the smallest THV area measured at post-procedural MSCT (p < 0.001). Moreover, reduced THV expansion was observed among patients with higher calcium burden (p = 0.048). Then, a new algorithm for TAVI sizing in BAV was develop (CASPER: Calcium Algorithm Sizing for bicusPid Evaluation with Raphe). This algorithm is based on the reassessment of the perimeter/area derived annulus diameter, according to three main anatomical features: 1) the ratio between raphe length and annulus diameter; 2)calcium burden; 3)calcium distribution in relation to the raphe. The algorithm was then validated in a new cohort of 21 patients, achieving 100% of procedural success and excellent TAVI performance. Conclusion MSCT assessment of raphe length, calcium burden and its distribution is of crucial relevance in the pre-procedural evaluation of patients with BAV. These anatomical features can be combined in a new and simple algorithm for TAVI sizing.

Bicuspid aortic valve sizing for transcatheter aortic valve implantation: Development and validation of an algorithm based on multi-slice computed tomography

Anna S. Petronio;Marco Angelillis;
2020-01-01

Abstract

Background No indication are available for transcatheter aortic valve implantation (TAVI) sizing in bicuspid aortic valve (BAV). Aim of the study is to develop and validate a Multi-Slice Computed Tomography (MSCT)-based algorithm for transcatheter heart valve (THV) sizing in patients with stenotic BAV under evaluation for TAVI. Methods A two steps method was applied: 1)evaluation of a cohort of 19 consecutive patients with type I BAV stenosis undergoing TAVI through pre and post-procedural MSCT, and development of an algorithm for THV sizing; 2)validation of the algorithm on a new cohort of 21 patients. Results In the first cohort, a high correlation was found between the raphe-level area measured at pre-procedural MSCT and the smallest THV area measured at post-procedural MSCT (p < 0.001). Moreover, reduced THV expansion was observed among patients with higher calcium burden (p = 0.048). Then, a new algorithm for TAVI sizing in BAV was develop (CASPER: Calcium Algorithm Sizing for bicusPid Evaluation with Raphe). This algorithm is based on the reassessment of the perimeter/area derived annulus diameter, according to three main anatomical features: 1) the ratio between raphe length and annulus diameter; 2)calcium burden; 3)calcium distribution in relation to the raphe. The algorithm was then validated in a new cohort of 21 patients, achieving 100% of procedural success and excellent TAVI performance. Conclusion MSCT assessment of raphe length, calcium burden and its distribution is of crucial relevance in the pre-procedural evaluation of patients with BAV. These anatomical features can be combined in a new and simple algorithm for TAVI sizing.
2020
Petronio, Anna S.; Angelillis, Marco; De Backer, Ole; Giannini, Cristina; Costa, Giulia; Fiorina, Claudia; Castriota, Fausto; Bedogni, Francesco; Labo...espandi
File in questo prodotto:
File Dimensione Formato  
Bicuspid aortic valve sizing.pdf

non disponibili

Tipologia: Versione finale editoriale
Licenza: NON PUBBLICO - accesso privato/ristretto
Dimensione 1.12 MB
Formato Adobe PDF
1.12 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/1146542
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 43
  • ???jsp.display-item.citation.isi??? 45
social impact