Aims: Permanent pacemaker implantation (PPM-I) remains nowadays the most important drawback of transcatheter aortic valve replacement (TAVR) procedure and the optimal strategy of delayed conduction disturbances (CDs) in these patients is unclear. The study aimed to validate an ambulatory electrocardiogram (ECG) monitoring through a 30 s spot ambulatory digital mobile ECG (AeECG), by using KardiaMobile-6L device in a 30-day period after TAVR procedure. Methods and results: Between March 2021 and February 2022, we consecutively enrolled all patients undergoing TAVR procedure, except pacemaker (PM) carriers. At discharge, all patients were provided of a KardiaMobile-6L device and a spot digital ECG (eECG) recording 1 month schedule. Clinical and follow-up data were collected, and eECG schedule compliance and recording quality were explored. Among 151 patients without pre-existing PM, 23 were excluded for pre-discharge PPM-I, 18 failed the KardiaMobile-6L training phase, and 10 refused the device. Delayed CDs with a Class I/IIa indication for PPM-I occurred in eight patients (median 6 days). Delayed PPM-I vs. non-delayed PPM-I patients were more likely to have longer PR and QRS intervals at discharge. PR interval at discharge was the only independent predictor for delayed PPM-I at multivariate analysis. The overall eECG schedule compliance was 96.5%. None clinical adverse events CDs related were documented using this new AeECG monitoring modality. Conclusion: A strategy of 30 s spot AeECG is safe and efficacious in delayed CDs monitoring after TAVR procedure with a very high eECG schedule level of compliance.

Non-continuous mobile electrocardiogram monitoring for post-transcatheter aortic valve replacement delayed conduction disorders put to the test

Parollo, Matteo;Barletta, Valentina;Giannotti Santoro, Mario;Primerano, Chiara;Angelillis, Marco;De Carlo, Marco;Zucchelli, Giulio;Bongiorni, Maria Grazia
Penultimo
;
Petronio, Anna Sonia
Ultimo
2023-01-01

Abstract

Aims: Permanent pacemaker implantation (PPM-I) remains nowadays the most important drawback of transcatheter aortic valve replacement (TAVR) procedure and the optimal strategy of delayed conduction disturbances (CDs) in these patients is unclear. The study aimed to validate an ambulatory electrocardiogram (ECG) monitoring through a 30 s spot ambulatory digital mobile ECG (AeECG), by using KardiaMobile-6L device in a 30-day period after TAVR procedure. Methods and results: Between March 2021 and February 2022, we consecutively enrolled all patients undergoing TAVR procedure, except pacemaker (PM) carriers. At discharge, all patients were provided of a KardiaMobile-6L device and a spot digital ECG (eECG) recording 1 month schedule. Clinical and follow-up data were collected, and eECG schedule compliance and recording quality were explored. Among 151 patients without pre-existing PM, 23 were excluded for pre-discharge PPM-I, 18 failed the KardiaMobile-6L training phase, and 10 refused the device. Delayed CDs with a Class I/IIa indication for PPM-I occurred in eight patients (median 6 days). Delayed PPM-I vs. non-delayed PPM-I patients were more likely to have longer PR and QRS intervals at discharge. PR interval at discharge was the only independent predictor for delayed PPM-I at multivariate analysis. The overall eECG schedule compliance was 96.5%. None clinical adverse events CDs related were documented using this new AeECG monitoring modality. Conclusion: A strategy of 30 s spot AeECG is safe and efficacious in delayed CDs monitoring after TAVR procedure with a very high eECG schedule level of compliance.
2023
De Lucia, Raffaele; Giannini, Cristina; Parollo, Matteo; Barletta, Valentina; Costa, Giulia; Giannotti Santoro, Mario; Primerano, Chiara; Angelillis, ...espandi
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/1165549
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact