Range verification with in-beam PET techniques is a powerful tool for monitoring the correctness of dose delivery in particle therapy. Among the major limitations of in-beam PET systems are the limited detectors size due to the constrained space in which they can be placed to allow the irradiation, and the necessity of a high read-out modularization, due to high activity rates during the irradiation. In this work we present the data acquired at the CNAO (Centro Nazionale di Adroterapia Oncologica) treatment center in Pavia, Italy, with the new DoPET system, specifically designed to operate in in-beam conditions. The new prototype consists of two planar 15cm x 15cm LYSO-based detectors, read out by 9 PMT detector modules each. In particular, we test the capability of our system to determine particle range in various irradiation conditions. Several plastic phantoms were irradiated at the CNAO treatment centre with protons and carbon ions of various energies. The used dose in treatment plans is 2 Gy and the monitoring feedback is produced in a few minutes after the end of the treatment.

In-beam PET data characterization with the large area DoPET prototype

SPORTELLI, GIANCARLO;BELCARI, NICOLA;CAMARLINGHI, NICCOLO';ZACCARO, EMANUELE;DEL GUERRA, ALBERTO;ROSSO, VALERIA
2016-01-01

Abstract

Range verification with in-beam PET techniques is a powerful tool for monitoring the correctness of dose delivery in particle therapy. Among the major limitations of in-beam PET systems are the limited detectors size due to the constrained space in which they can be placed to allow the irradiation, and the necessity of a high read-out modularization, due to high activity rates during the irradiation. In this work we present the data acquired at the CNAO (Centro Nazionale di Adroterapia Oncologica) treatment center in Pavia, Italy, with the new DoPET system, specifically designed to operate in in-beam conditions. The new prototype consists of two planar 15cm x 15cm LYSO-based detectors, read out by 9 PMT detector modules each. In particular, we test the capability of our system to determine particle range in various irradiation conditions. Several plastic phantoms were irradiated at the CNAO treatment centre with protons and carbon ions of various energies. The used dose in treatment plans is 2 Gy and the monitoring feedback is produced in a few minutes after the end of the treatment.
2016
Sportelli, Giancarlo; Belcari, Nicola; Camarlinghi, Niccolo'; Ciocca, M.; Collini, F.; Molinelli, S.; Pullia, M.; Zaccaro, Emanuele; DEL GUERRA, Alberto; Rosso, Valeria
File in questo prodotto:
File Dimensione Formato  
jinst16_02_c02089.pdf

solo utenti autorizzati

Tipologia: Versione finale editoriale
Licenza: Importato da Ugov Ricerca - Accesso privato/ristretto
Dimensione 2.05 MB
Formato Adobe PDF
2.05 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/819459
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact