Objectives: We report procedural and 30-day clinical outcomes following transcatheter aortic valve replacement (TAVR) with Lotus Valve system in a high-risk population. Background: Lotus valve is a second-generation TAVR fully repositionable and retrievable device. RELEVANT (REgistry of Lotus valvE for treatment of aortic VAlve steNosis with Tavr) study is an Italian prospective multicentre registry. Methods: Five major centers performing TAVR using Lotus participated. All high-risk symptomatic patients with severe aortic stenosis were evaluated by a heart team and screened for eligibility for TAVR. Primary end-points were procedural and 30-day mortality. Secondary endpoints included procedural and 30-day safety/effectiveness metrics according to Valve Academic Research Consortium (VARC)-2 criteria. Results: Two hundred and twenty-five patients undergoing TAVR with Lotus were enrolled. Mean age was 82.6±6.3 years, 51.6% females. Mean STS score for mortality was 8.3±5.6. Procedural success was achieved in 98.7% of patients. All-cause mortality was 2.2% at discharge and 2.7% at 30-day. Stroke rate was 2.2% at discharge and 3.1% at 30-day. Patients requiring a new permanent pacemaker implantation were 30.7% at discharge and 31.8% at 30-day. Paravalvular regurgitation (PVR) was trace/mild in 99.1% of patients at discharge. Only two patients (0.9%) had moderate PVR at discharge and at 30-day, whereas none had severe PVR. Conclusions: RELEVANT study showed that TAVR using Lotus, in a real-world population of patients with severe aortic stenosis at high surgical risk, was associated with excellent device success implantation and early safety according to VARC-2 definition. The pacemaker implantation rate was about a third of patients.

Procedural and 30-day clinical outcomes following transcatheter aortic valve replacement with lotus valve: Results of the RELEVANT study

ANGELILLIS, MARCO;PETRONIO, ANNA;
2017-01-01

Abstract

Objectives: We report procedural and 30-day clinical outcomes following transcatheter aortic valve replacement (TAVR) with Lotus Valve system in a high-risk population. Background: Lotus valve is a second-generation TAVR fully repositionable and retrievable device. RELEVANT (REgistry of Lotus valvE for treatment of aortic VAlve steNosis with Tavr) study is an Italian prospective multicentre registry. Methods: Five major centers performing TAVR using Lotus participated. All high-risk symptomatic patients with severe aortic stenosis were evaluated by a heart team and screened for eligibility for TAVR. Primary end-points were procedural and 30-day mortality. Secondary endpoints included procedural and 30-day safety/effectiveness metrics according to Valve Academic Research Consortium (VARC)-2 criteria. Results: Two hundred and twenty-five patients undergoing TAVR with Lotus were enrolled. Mean age was 82.6±6.3 years, 51.6% females. Mean STS score for mortality was 8.3±5.6. Procedural success was achieved in 98.7% of patients. All-cause mortality was 2.2% at discharge and 2.7% at 30-day. Stroke rate was 2.2% at discharge and 3.1% at 30-day. Patients requiring a new permanent pacemaker implantation were 30.7% at discharge and 31.8% at 30-day. Paravalvular regurgitation (PVR) was trace/mild in 99.1% of patients at discharge. Only two patients (0.9%) had moderate PVR at discharge and at 30-day, whereas none had severe PVR. Conclusions: RELEVANT study showed that TAVR using Lotus, in a real-world population of patients with severe aortic stenosis at high surgical risk, was associated with excellent device success implantation and early safety according to VARC-2 definition. The pacemaker implantation rate was about a third of patients.
2017
Montone, Rocco A; Testa, Luca; Fraccaro, Chiara; Montorfano, Matteo; Castriota, Fausto; Nerla, Roberto; Angelillis, Marco; Tusa, Maurizio; Giannini, Francesco; Garatti, Andrea; Tarantini, Giuseppe; Petronio, Anna; Brambilla, Nedy; Bedogni, Francesco
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/847850
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 12
  • ???jsp.display-item.citation.isi??? ND
social impact