A reliable method for the determination of carbonyl compounds in exhaled breath based on on-sorbent derivatization coupled with thermal desorption and gas chromatography-tandem mass spectrometry is described. The analytical performances were optimized for a mixture of C2-C9 aldehydes and C3-C9 ketones, particularly interesting for clinical applications, by using an internal standard and applying a 2^3 full factorial design. A volume of sample (250 ml) was loaded at 50 ml min-1 into a Tenax GR sorbent tube containing 130 nmol of O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine hydrochloride. All compounds showed a limit of detection lower than 200 pptv. The yield of the derivatization procedure was normalized by adding to the sample a known amount of 6D-acetone as an internal standard. This allowed halving the relative standard deviation to 10% and 15% for the mono-and di-carbonyl compounds, respectively, thus improving reliability. The optimized method was applied to the determination of carbonyl compounds in 12 breath samples collected from four patients suffering from heart failure during hospitalization.

Determination of carbonyl compounds in exhaled breath by on-sorbent derivatization coupled with thermal desorption and gas chromatography-tandem mass spectrometry

Lomonaco, T
Primo
;
Ghimenti, S;Biagini, D;Salvo, P;Fuoco, R;Di Francesco, F
2018-01-01

Abstract

A reliable method for the determination of carbonyl compounds in exhaled breath based on on-sorbent derivatization coupled with thermal desorption and gas chromatography-tandem mass spectrometry is described. The analytical performances were optimized for a mixture of C2-C9 aldehydes and C3-C9 ketones, particularly interesting for clinical applications, by using an internal standard and applying a 2^3 full factorial design. A volume of sample (250 ml) was loaded at 50 ml min-1 into a Tenax GR sorbent tube containing 130 nmol of O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine hydrochloride. All compounds showed a limit of detection lower than 200 pptv. The yield of the derivatization procedure was normalized by adding to the sample a known amount of 6D-acetone as an internal standard. This allowed halving the relative standard deviation to 10% and 15% for the mono-and di-carbonyl compounds, respectively, thus improving reliability. The optimized method was applied to the determination of carbonyl compounds in 12 breath samples collected from four patients suffering from heart failure during hospitalization.
2018
Lomonaco, T; Romani, A; Ghimenti, S; Biagini, D; Bellagambi, F G; Onor, M; Salvo, P; Fuoco, R; Di Francesco, F
File in questo prodotto:
File Dimensione Formato  
Determination of carbonyl compounds.pdf

solo utenti autorizzati

Descrizione: Articolo finale
Tipologia: Versione finale editoriale
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 1.59 MB
Formato Adobe PDF
1.59 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
JBR 100799 Rev.pdf

Open Access dal 01/01/2020

Tipologia: Documento in Post-print
Licenza: Creative commons
Dimensione 530.98 kB
Formato Adobe PDF
530.98 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11568/926043
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 18
social impact