We present an autoradiography tool to be used mainly for genetic studies. It performs a quantitative analysis of radioactivity and can follow a dynamic process. We designed several applications, in particular one aimed at detecting hybridization of radio-labeled DNA fragments with known DNA-probes deposited on a micro-array. The technique is based on GaAs pixel array detector and low threshold, large dynamic range and good sensitivity integrated electronics developed for medical applications, suitable to detect markers (gamma or beta) such as C-14, S-35, P-33, P-32, I-125, even at very low activities. A Monte Carlo simulation of beta(-) detection in GaAs is presented here in order to study the spatial resolution characteristics of such a system. For several biological applications, the electronics is required to perform at high temperatures (from 37 degrees to 68 degrees): we present here studies of noise and minimum threshold as a function of the temperature. (C) 1999 Elsevier Science B.V. All rights reserved. RI maestro, paolo/E-3280-2010
GaAs pixel radiation detector as an autoradiography tool for genetic studies
BISOGNI, MARIA GIUSEPPINA;Ciocci MA;DELOGU, PASQUALE;FANTACCI, MARIA EVELINA;ROSSO, VALERIA;
1999-01-01
Abstract
We present an autoradiography tool to be used mainly for genetic studies. It performs a quantitative analysis of radioactivity and can follow a dynamic process. We designed several applications, in particular one aimed at detecting hybridization of radio-labeled DNA fragments with known DNA-probes deposited on a micro-array. The technique is based on GaAs pixel array detector and low threshold, large dynamic range and good sensitivity integrated electronics developed for medical applications, suitable to detect markers (gamma or beta) such as C-14, S-35, P-33, P-32, I-125, even at very low activities. A Monte Carlo simulation of beta(-) detection in GaAs is presented here in order to study the spatial resolution characteristics of such a system. For several biological applications, the electronics is required to perform at high temperatures (from 37 degrees to 68 degrees): we present here studies of noise and minimum threshold as a function of the temperature. (C) 1999 Elsevier Science B.V. All rights reserved. RI maestro, paolo/E-3280-2010I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.