Whole-Room Indirect Calorimeters (WRIC) are accurate tools to precisely measure energy metabolism in humans via calculation of oxygen consumption and carbon dioxide production. Yet, overall accuracy of metabolic measurements relies on the validity of the dynamic model for gas exchange inside the calorimeter volume in addition to experimental and environmental conditions that contribute to the uncertainty of WRIC outcome variables. The aim of this work is to formally study the sensitivity of a WRIC system operated in a push configuration at the steady-state condition to identify the optimal experimental conditions to obtain the best degree of accuracy for outcome metabolic measurements. The results of our sensitivity analysis are then validated with measurements obtained during propane combustion tests performed at the WRIC located at the University Hospital of Pisa. Our results demonstrate that achieving a fractional concentration of carbon dioxide inside the calorimeter >0.2% leads to relative uncertainty <5% for the outcome metabolic measurements when assuming an accuracy class of 1% for gas analyzer instruments.
Sensitivity Analysis of Whole-Room Indirect Calorimeters at the Steady-state Condition
Bandini, GabrielePrimo
;Landi, Alberto;Santini, Ferruccio;Basolo, Alessio;Marracci, Mirko;Piaggi, Paolo
Ultimo
2023-01-01
Abstract
Whole-Room Indirect Calorimeters (WRIC) are accurate tools to precisely measure energy metabolism in humans via calculation of oxygen consumption and carbon dioxide production. Yet, overall accuracy of metabolic measurements relies on the validity of the dynamic model for gas exchange inside the calorimeter volume in addition to experimental and environmental conditions that contribute to the uncertainty of WRIC outcome variables. The aim of this work is to formally study the sensitivity of a WRIC system operated in a push configuration at the steady-state condition to identify the optimal experimental conditions to obtain the best degree of accuracy for outcome metabolic measurements. The results of our sensitivity analysis are then validated with measurements obtained during propane combustion tests performed at the WRIC located at the University Hospital of Pisa. Our results demonstrate that achieving a fractional concentration of carbon dioxide inside the calorimeter >0.2% leads to relative uncertainty <5% for the outcome metabolic measurements when assuming an accuracy class of 1% for gas analyzer instruments.File | Dimensione | Formato | |
---|---|---|---|
TIM_MEMEA_WRICSTATICSENS_final.pdf
accesso aperto
Tipologia:
Documento in Post-print
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
1.08 MB
Formato
Adobe PDF
|
1.08 MB | Adobe PDF | Visualizza/Apri |
Bandini (2023 TIM) static sensitivity chamber.pdf
non disponibili
Tipologia:
Versione finale editoriale
Licenza:
NON PUBBLICO - accesso privato/ristretto
Dimensione
5.71 MB
Formato
Adobe PDF
|
5.71 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.